Journal of Medical Physics
ORIGINAL ARTICLE
Year
: 2011  |  Volume : 36  |  Issue : 1  |  Page : 3--14

Small field dose delivery evaluations using cone beam optical computed tomography-based polymer gel dosimetry


Timothy Olding1, Oliver Holmes1, Paul DeJean1, Kim B McAuley2, Ken Nkongchu3, Giles Santyr4, L John Schreiner5 
1 Department of Physics, Queen's University, Kingston, Ontario, Canada
2 Department of Chemical Engineering, Queen's University, Kingston, Ontario, Canada
3 Henry Ford Hospital, Detroit, MI, USA
4 Imaging Research Laboratories, Robarts Research Institute, London, ON; Department of Medical Biophysics and Medical Imaging, University of Western Ontario, London, ON, Canada
5 Department of Physics, Queen's University, Kingston, Ontario; Department of Oncology, Queen's University, Kingston, Ontario; Department of Medical Physics, Cancer Centre of Southeastern Ontario at Kingston General Hospital, Kingston, Ontario, Canada

Correspondence Address:
Timothy Olding
Department of Medical Physics, Cancer Centre of Southeastern Ontario at Kingston General Hospital, 25 King Street West, Kingston, Ontario K7L 5P9
Canada

This paper explores the combination of cone beam optical computed tomography with an N-isopropylacrylamide (NIPAM)-based polymer gel dosimeter for three-dimensional dose imaging of small field deliveries. Initial investigations indicate that cone beam optical imaging of polymer gels is complicated by scattered stray light perturbation. This can lead to significant dosimetry failures in comparison to dose readout by magnetic resonance imaging (MRI). For example, only 60% of the voxels from an optical CT dose readout of a 1 l dosimeter passed a two-dimensional Low«SQ»s gamma test (at a 3%, 3 mm criteria, relative to a treatment plan for a well-characterized pencil beam delivery). When the same dosimeter was probed by MRI, a 93% pass rate was observed. The optical dose measurement was improved after modifications to the dosimeter preparation, matching its performance with the imaging capabilities of the scanner. With the new dosimeter preparation, 99.7% of the optical CT voxels passed a Low«SQ»s gamma test at the 3%, 3 mm criteria and 92.7% at a 2%, 2 mm criteria. The fitted interjar dose responses of a small sample set of modified dosimeters prepared (a) from the same gel batch and (b) from different gel batches prepared on the same day were found to be in agreement to within 3.6% and 3.8%, respectively, over the full dose range. Without drawing any statistical conclusions, this experiment gives a preliminary indication that intrabatch or interbatch NIPAM dosimeters prepared on the same day should be suitable for dose sensitivity calibration.


How to cite this article:
Olding T, Holmes O, DeJean P, McAuley KB, Nkongchu K, Santyr G, Schreiner L J. Small field dose delivery evaluations using cone beam optical computed tomography-based polymer gel dosimetry.J Med Phys 2011;36:3-14


How to cite this URL:
Olding T, Holmes O, DeJean P, McAuley KB, Nkongchu K, Santyr G, Schreiner L J. Small field dose delivery evaluations using cone beam optical computed tomography-based polymer gel dosimetry. J Med Phys [serial online] 2011 [cited 2022 Aug 10 ];36:3-14
Available from: https://www.jmp.org.in/article.asp?issn=0971-6203;year=2011;volume=36;issue=1;spage=3;epage=14;aulast=Olding;type=0