On-line Beam Data Acquisition System And Beam Data Analysis


Paul B Ravindran1, C A Jayachandran2 
1 
2 

Correspondence Address:
Paul B Ravindran


An indigenous microprocessor controlled Radiation Field Analyzer has been in use for the last one year in this hospital. It has completely replaced the laborious manual procedure of beam data collection. This paper describes the improvements carried out in the mechanical system for the detector movement, the on-line data acquisition and beam data analysis software developed for an IBM compatible personal computer and the results of the measurements carried out on teletherapy units. Beam profiles and central axis depth dose measurements for open, wedge and shaped beams carried out on teletherapy units and the isodose curves generated with these data are also presented in this paper.


How to cite this article:
Ravindran P, Jayachandran C. On-line Beam Data Acquisition System And Beam Data Analysis.J Med Phys 1995;20:9-13


How to cite this URL:
Ravindran P, Jayachandran C. On-line Beam Data Acquisition System And Beam Data Analysis. J Med Phys [serial online] 1995 [cited 2022 May 25 ];20:9-13
Available from: https://www.jmp.org.in/article.asp?issn=0971-6203;year=1995;volume=20;issue=2;spage=9;epage=13;aulast=Ravindran;type=0