Reduction of Metal Artifacts Caused by Titanium Peduncular Screws in the Spine by Means of Monoenergetic Images and the Metal Artifact Reduction Software in Dual-Energy Computed Tomography


1 Department of Diagnostic and Interventional Radiology, IRCCS Istituto Ortopedico Rizzoli, Via Albertoni, Bologna, Italy
2 Department of Radiology, IRCCS Azienda Ospedaliero-Universitaria Di Bologna, Via Albertoni, Bologna, Italy

Correspondence Address:
Dr. Giulio Vara
Via Ernesto Masi 43, 40137, Bologna
Italy
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmp.jmp_121_21

Rights and Permissions

Objectives: To evaluate the reduction of metal artifacts in patients with titanium peduncular screws in the spine using (1) conventional images (CI), (2) virtual monoenergetic reconstructions (VMRs), and (3) VMR + Metal Artifact Reduction Software (VMR + MARS), with dual-energy computed tomography (DECT). Materials and Methods: Twenty-four patients with titanium peduncular screws in the spine were studied using a 64-channel DECT. During the postprocessing phase, the CI, the VMRs from 100 to 140 keV, and the VMR at 140 keV + MARS were synthesized. All the images were considered, and a quantitative evaluation was performed measuring the attenuation values (in terms of Hounsfield Units) with region of interest, in correspondence with the most hyperdense and hypodense artifacts. All the values were then compared. A qualitative evaluation, in terms of image quality and extent of artifacts, was also performed by two radiologists. Results: In quantitative terms, the 140 keV + MARS reconstruction was able to significantly reduce both bright and dark metal artifacts, compared to CI and to VMRs. The VMR was capable of significantly reducing both dark and bright artifacts, compared to CI. In qualitative terms, the VMR at 140 keV proved to be the best, compared to CI and VMR + MARS images. Conclusions: The VMR + MARS image reduces metal artifacts from titanium peduncular screws more than VMRs alone and CI. Furthermore, the VMR can decrease metal artifacts from a quantitative and a qualitative point of view. Combining information from VMRs and VMR + MARS images could be the best way to solve the issue of metal artifacts on computed tomography images.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed142    
    Printed0    
    Emailed0    
    PDF Downloaded36    
    Comments [Add]    

Recommend this journal