ORIGINAL ARTICLE
Year : 2022  |  Volume : 47  |  Issue : 1  |  Page : 73-78

Evaluation of radiation dose reduction and its effect on image quality for different flat-panel detectors


Department of Radiology, Christian Medical College, Vellore, Tamil Nadu, India

Correspondence Address:
Dr. Roshan Samuel Livingstone
Department of Radiology, Christian Medical College, Vellore - 632 004, Tamil Nadu
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmp.jmp_127_21

Rights and Permissions

Purpose: To evaluate the image quality of semi-anatomical chest radiographs acquired using low radiation doses from seven different flat-panel detector (FPD) systems. Materials and Methods: Radiographs of a semi-anatomical chest phantom were acquired at 70 and 110 kVp using 7 different FPDs from 5 vendors. Radiation doses were measured using a dose-area-product (DAP) meter. To standardize measurements across all FPDs, DAP value of 51.05 μGym2 obtained at 70 kVp and 9.43 μGym2 at 110 kVp was used as reference in this study. Radiation doses were reduced by manually adjusting mAs for both tube potentials in all FPD systems to achieve acceptable image quality. Contrast-to-noise ratio, signal-to-noise ratio and figure of merit (FOM) in lung, heart, and diaphragm for all images were analyzed. Results: In comparison with set standard protocol, radiation dose reduction of 20%, 21%, 21.6%, 59.5%, 60.7%, 62.2%, and 67.6% with optimal image quality was observed in Prognosys Prorad, GE Definium 8000, Siemens Fusion, Fujifilm FGX, Fujifilm FGXR, Philips Digital Diagnost and Siemens Aristos at 70 kVp. At 110 kVp, dose reduction of 15.7% and 34.8% was possible only for Philips Digital Diagnost and Siemens Aristos. FOM was high at 110 kVp even when radiation doses were reduced by a factor 2 when compared to 70 kVp in all digital radiography systems. Conclusion: This study demonstrates the feasibility of using a semi-anatomical chest phantom in the optimization of radiation dose and image quality. The FOM was a good indicator in assessing image quality between different detectors.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed596    
    Printed32    
    Emailed0    
    PDF Downloaded76    
    Comments [Add]    

Recommend this journal