ORIGINAL ARTICLE
Year : 2022  |  Volume : 47  |  Issue : 1  |  Page : 65-72

Influence of jaw setting in the determination of stereotactic small-field output factors with different detectors


Department of Radiation Oncology, Christian Medical College, Vellore, Tamil Nadu, India

Correspondence Address:
Dr. A Sathish Kumar
Department of Radiation Oncology, Christian Medical College, Vellore - 632 004, Tamil Nadu
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmp.jmp_111_21

Rights and Permissions

Background: The experimental determination of relative output factors presents the greatest challenge, especially for small fields with different detectors. The aim of this study is to evaluate the influence of jaw positions on small-field output factors for the fields defined by micro-multileaf collimator and circular cones with different detectors. Materials and Methods: The stereotactic output factors were measured on Primus linear accelerator with BrainLab micro-multileaf collimator (mMLC) and circular cones as add-on tertiary collimators. Square field sizes ranging from 0.6 cm × 0.6 cm to 9.8 cm × 9.8 cm and circular fields of diameter ranging from 1.0 cm to 4.0 cm were defined by mMLC and circular cones, respectively. The influence of jaw position on output factor was assessed for different geometric configurations with three different detectors. Results: The values obtained with PinPoint ion chamber were consistent with microDiamond detector for fields greater than 24 mm × 24 mm, but an underestimation of 23.9% was noticed in 6 mm x 6 mm field size. For the mMLC defined field size of 6 mm × 6 mm, when the X-Y jaw was moved from 8 mm × 8 mm to 80 mm × 80 mm, an increase in the output by a factor of 1.7 was observed with both microDiamond and stereotactic radiosurgery diode, whereas an increase in output by a factor of 1.9 was noticed with PinPoint ion chamber. Conclusion: Output factors obtained with different detectors show high differences in the smallest field size for all collimating systems. This study confirms that the position of X and Y jaw above the tertiary collimator significantly influences the small-field output factor.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed160    
    Printed6    
    Emailed0    
    PDF Downloaded30    
    Comments [Add]    

Recommend this journal