ORIGINAL ARTICLE
Year : 2022  |  Volume : 47  |  Issue : 1  |  Page : 20-26
 

Impact of dosimetric parameters on interplay effects in 6 MV flattening filter-free photon beams to treat lung cancer


1 Medical Physics Program, Department of Radiology, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand
2 Division of Radiation Oncology, Department of Radiology, King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand

Date of Submission01-May-2021
Date of Decision15-Jan-2022
Date of Acceptance15-Jan-2022
Date of Web Publication31-Mar-2022

Correspondence Address:
Dr. Taweap Sanghangthum
Department of Radiology, Division of Radiation Oncology, Faculty of Medicine, Chulalongkorn University, Bangkok
Thailand
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmp.jmp_63_21

Rights and Permissions

 

   Abstract 

Context: Interplay effects have become the significant problem in lung cancer radiotherapy. Since these effects yield dose variation within the target and surrounding tissues. Aim: The aim of this study is to investigate the effect of the dosimetric parameters of interplay effects in 6 MV flattening filter-free (FFF) photon beams for lung cancer. Settings and Design: This study performed planning, measurement, and data analysis sections for examining different breathing amplitudes and phases, doses, dose rates, field sizes, and fractionations. Subjects and Methods: Standard and clinical plans were created on the eclipse treatment planning system. The static and dynamic measurements were performed using a robotic platform and two-dimensional (2D) diode array. The gamma passing rates were defined as the percent of dose variation caused by the interplay effects. Statistical Analysis Used: Unpaired t-test. Results: The outcomes showed three trends between gamma passing rates (γ) and dosimetric parameters. First, a decreasing trend was breathing amplitudes. The lowest γ of maximum amplitudes (2 cm) in both one dimensional and 2D were <25%. Second, an increasing trend was field sizes. The lowest γ of minimum field size (4 cm × 4 cm2) was <55%. Third, constant outcomes were breathing phases, doses, dose rates, and a number of fractions. The γ values of these factors were 53.1%, 55.1%, 34.7%, and 36.7%, respectively. Conclusions: Lung tumor motion-induced interplay effects in 6 MV FFF photon beams are more pronounced for higher breathing amplitudes and smaller field sizes.


Keywords: Flattening filter free, interplay effects, lung cancer


How to cite this article:
Poolnapol V, Sanghangthum T, Keawsamur M. Impact of dosimetric parameters on interplay effects in 6 MV flattening filter-free photon beams to treat lung cancer. J Med Phys 2022;47:20-6

How to cite this URL:
Poolnapol V, Sanghangthum T, Keawsamur M. Impact of dosimetric parameters on interplay effects in 6 MV flattening filter-free photon beams to treat lung cancer. J Med Phys [serial online] 2022 [cited 2022 May 22];47:20-6. Available from: https://www.jmp.org.in/text.asp?2022/47/1/20/341423



   Introduction Top


Lung cancer is the most common malignancy in both genders worldwide. It is the leading cause of death, approximately 2.09 million cases in 2018.[1] Lung stereotactic body radiation therapy (SBRT) treatment has been widely used in recent years.[2],[3],[4] SBRT is the treatment technique that delivers extremely precise intense doses of radiation to cancer cells while minimizing damage to healthy tissue. Furthermore, the SBRT also provides large doses per few fractions together with a rapid fall-off of dose outside the target. Hence, the accuracy of dose delivery and conformity of dose is very essential. However, this technique still poses a challenge to the treatment of lung cancer. The most important concern is interplay effects due to patient breathing during irradiation. This interplay effect, the effects between tumor motion and multileaf collimators (MLCs) motion occurring at the same time, leads to heterogeneities within the target volume and/or unwanted dose to the surrounding tissue. The previous study of Mukhlisin et al. indicated that tumor motion during irradiation can bring dose validation inside the tumor target in both lung intensity-modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) treatment techniques.[5] Likewise, Adamczyk et al. study illustrated that lung three-dimensional conformal radiotherapy (3D-CRT) and IMRT treatment techniques could cause tumor underdosage due to the interplay effect.[6] In the same way, Kubo et al. showed the number of patient breathing induced by interplay effects also impact the dose variation in lung VMAT-SBRT.[7] Moreover, several studies revealed the blurring of dose distribution of tumor target during irradiation owing to the interplay effects when tumor was motioned in only craniocaudal (CC) direction (one dimensional [1D] motion).[8],[9],[10],[11] However, in recent years, flattening filter-free (FFF) technique was utilized together with the SBRT technique to increase the efficiency of lung cancer treatment[12],[13] and reduce the interplay effects.[9] It has several benefits such as providing a higher dose rate with shorter time treatment as well as allowing inhomogeneous dose distribution with reduced peripheral dose.[14],[15],[16] As a result, the goal of this study is to investigate the dosimetric parameters of interplay effects in 6 MV FFF photon beams with different breathing amplitudes and phases, doses, dose rates, field sizes, and fraction numbers using two-dimensional (2D) robotic platform together with the 2D diode array.


   Subjects and Methods Top


Robotic platform: The MotionSimXY/four-dimensional (4D) platform from Sun Nuclear Corporation, Melbourne, FL was a precision instrument designed to use together with the 2D diode array for quality assurance study of motion effects in radiation therapy imaging and delivery by moving a phantom through programmable motion patterns (an operating system same as MapCHECK2). The options of the phantom were provided accuracy and precision of X-Y motor drive to move the MapCHECK2 in X and Y directions. Moreover, this platform also allowed a wide range of motor movement, which the maximum travel was 10.2 cm of each X-Y axis and the maximum velocity was 5.08 cm/s.

Robotic platform: The MapCHECK2 (Sun Nuclear Corporation, Melbourne, FL) is a 2D detector array for accurate and fast verifying planar radiotherapy dose distributions. It offers smaller 1,527 diode detectors placed uniformly throughout the array provide high sensitivity and proven stability in a large active field size 26 cm × 32 cm2. A real-time electrometer measures every pulse with 50 ms updates.

Patient information: To test the interplay effects impact of the clinical plans, a retrospective group of eight patients with lung cancer who underwent 4D computed tomography (4DCT) at free-breathing conditions and 2–3 full arcs of SBRT-VMAT during 2017–2020 was chosen. The maximum intensity projection reconstruction image from ten phases of 4DCT images was transferred to Eclipse treatment planning system (TPS) (Varian Medical System, Palo Alto, CA) to be set as the planning image, where the irradiation targets were the planning target volume (PTV) that expanded margin from the clinical target volume of 0.5–1.0 cm. The range of total PTV volumes was 9.1–83.0 cm3. The number of PTV was 1–6, which averages located on the middle and upper lobes of both sides of the lung. For organs at risk (OARs), the OARs were heart, esophagus, spinal cord, whole lungs (subtract PTV), and contralateral lung.

Planning preparation

This study investigated the interplay effects from two groups of the plans. There were the standard and the clinical plans groups; each group was defined as follows:

Standard plans group

The 6 MV FFF of photon beams plans were created on Eclipse TPS version 15.6.05 with an Anisotropic Analytical Algorithm. These standard plans were planned with anteroposterior field with the planning parameters of 500 cGy/fraction prescription dose, 1400 MU/min dose rate, 4 cm × 4 cm2 field size, 100 cm source to axis distance, and 5 cm measurement depth.

Clinical plans group

The eight lung cancer plans were divided into two subgroups. There were four complex plans and four noncomplex plans. The complex plans refer to the plans with multiple targets. All clinical plans were optimized and calculated for VMAT partial arcs with a gantry angle from 70° to 290° to avoid the beams passing through MotionSimXY/4D platform. These eight clinical plans were also applied for 6 MV FFF photon energy with 1400 MU/min dose rate. The other planning parameters were similar to the original patient planning parameters such as the prescribed dose. The plans were evaluated by isodose distribution and dose-volume histogram. The standardized prescription isodose was chosen according to the following criteria: Ninety-five percentage of PTV was encompassed by the prescription isodose volume and 99% of PTV was covered by 95% of the prescription dose.[17] Finally, all plans were created into verification plans or QA plans for dose measurements. The prescription dose of each individual case was defined by a radiation oncologist. The range of total dose was 3,000–5,070 cGy in 3–6 fractions.

Dose measurement

The static and dynamic dose measurements were simulated on a robotic platform to examine the effects of different amplitudes and phases of the platform, which can represent the breathing pattern of the lung cancer patient. Moreover, this study also investigated on different doses, dose rates, field sizes, and number of fractions as illustrated in [Table 1]. All plans were irradiated on the robotic platform from TrueBeam linear accelerator (Varian Medical Systems, Palo Alto, CA, USA). The dose outcomes were measured by a 2D diode array positioned on the robotic platform and covered with 3 cm of the solid water phantom as shown in [Figure 1]. All measurements stipulated the robotic platform moving pattern in 1 cm of amplitude in the Y-axis and 4 s of phase that estimated from our patient breathing information. This motion was set according to these parameters except the amplitudes effect study because we have to vary the amplitudes to see the interplay effects outcomes. However, the phase still used at 4 s. The breathing velocity in all study was set to 0.5 cm/s.
Table 1: Robotic platform and three-dimensional conformal radiotherapy planning parameters setting

Click here to view
Figure 1: The setting up of robotic platform, MapCHECK2, and solid water phantom

Click here to view


Data analysis

The results of calculated and measured doses were compared using gamma analysis with 1%/1 mm criteria for standard plans. On the other hand, 3%/2 mm of criteria following the American Association of Physicists in Medicine Task Group number 218 recommendation was applied in the clinical part.[18] As for statistical analysis, an unpaired t-test was selected for checking of independent samples tests whether the mean values of two independent groups differed significantly. When the P value of the test is <0.05, it is assumed that there is a statistically significant difference in the variances between the two groups.


   Results Top


The results of this research were presented in the following experimental steps: Standard plans dose measurement and clinical plans dose measurement, respectively.

Standard plans dose measurement

[Table 2] displays a decreasing tendency between the gamma passing rates and amplitudes. The results show higher amplitudes received lower gamma passing rates in all moving patterns, especially for 2D motion (XY-axis). Nevertheless, these comparisons were not statistically significant differences in the results between 1D and 2D motion since all P values exceeded 0.05. The correlation between gamma passing rates and amplitudes was explained by a mathematical linear equation in form of negative direct variation as shown in [Figure 2]. The r-squared (R2) of the three curves provided more than 0.90, which can confirm the good negative agreements between gamma passing rate and motion in different amplitude. [Figure 3] exhibits an increasing tendency of gamma passing rates with different symmetric field sizes. The correlation curve illustrates the positive direct variation of the mathematical linear equation together with approximately R2 of 0.92. Other parameters of phase, dose, dose rate, and number of fractions revealed a constant trend of the results as shown in [Table 3].
Table 2: The gamma passing rates at 1%/1 mm criteria of standard plans irradiation with different amplitudes and unpaired t-test statistical analysis between 1D (X or Y-axis) and 2D (XY-axis) moving patterns

Click here to view
Figure 2: The correlation curves of all moving patterns between the gamma passing rates and the different amplitudes: (a) X-axis, (b) Y-axis, and (c) XY-axis

Click here to view
Figure 3: The correlation curve between the gamma passing rates and the different field sizes

Click here to view
Table 3: The gamma passing rates from 1%/1 mm criteria and the linear equations of standard plans irradiation with different phases, doses, dose rates, and number of fractions

Click here to view


Clinical plans dose measurement

[Table 4] shows a reduction of gamma passing rates when the platform was moved. The noncomplex plans group received more than 84.0% of gamma passing rates whereas the complex plans group acquired <80.0%. Furthermore, both groups expressed dose blurring and changing in the shape of dose distribution as displayed in [Figure 4] and [Figure 5] for noncomplex and complex plan, respectively.
Table 4: The gamma passing rates for 3%/2 mm criteria and the planning target volume information of eight clinical plans irradiation to the static and dynamic mode of robotic platform motion

Click here to view
Figure 4: Examples of dose distributions of noncomplex lung stereotactic body radiation therapy plan delivered to MapCHECK2 placing on the robotic platform: (a) Dose distribution during dynamic platform measurement and (b) Dose distribution from the planning step

Click here to view
Figure 5: Examples of dose distributions of complex lung stereotactic body radiation therapy plan delivered to MapCHECK2 placing on the robotic platform: (a) Dose distribution during dynamic platform measurement and (b) Dose distribution from the planning step

Click here to view



   Discussion Top


In this study, the percent of dose variation caused by the interplay effects were defined as the percent of the gamma passing rates. Thus, increasing the percent of gamma passing rates can be interpreted as a reduction of the impact of dosimetric parameters, which leads to an interplay effect. The results showed three tendencies of the standard plans dose measurement. First, the decrement trend was shown in amplitude outcomes. The results are in accordance with previous experimental research by Mukhlisin et al., Adamczyk et al., Kakakhel et al., and Edvardsson et al., which stated that the interplay effects were larger for higher breathing amplitudes of CC direction.[5],[6],[8],[9] Besides, according to [Figure 2], the slope value of the XY-amplitude correlation curve of this study showed higher the R2 value more than the R2 value of both X-amplitude and Y-amplitude slopes, whereas these two curves still receive the resemble outcomes. This point indicated that the 2D amplitudes motion has more impact on the interplay effects over 1D amplitude motions. Second, the interplay effects were reduced by the larger field size. By cause of the larger expanding area allows more doses matching accuracy between calculated and measured doses. Kakakhel et al. also observed the increased gamma passing rates depended on the larger width of the field size.[8] Therefore, all findings mentioned above revealed that the longer distance and the larger number of directions of the tumor motion together with smaller treatment field size increase the interplay effects. Because the larger displacement leads to the larger area where the radiation doses do not overlap, furthermore, these problems easily provide tumor to escape from the beam trajectory.[6] Thereby, precise patient breathing and treatment field verification are the most important factor to ensure correct planning and delivery for lung cancer treatment. Third, the parameters of phases, doses, dose rates, and number of fractions were not affected by the interplay effects when compared dynamic phantom to the static phantom (phase was zero). All the static measurements showed 100% of gamma passing rate because the simple 3D-CRT planning technique was performed in this research. This technique provides a homogeneous field without any unchanging beam intensity. Furthermore, for number of fractions study, the phantom moving pattern and the homogeneous plans were similar in every fraction. Therefore, the results do not change. The other perspectives from Court et al., Ong et al., Rao et al., and Adamczyk et al. presented that the decrease in dose variation was associated with the decreasing in dose rate.[19],[20],[21],[22],[23] The 3D-CRT plans were generated with the dose rate of 300 MU/min, which was less than twice dose rate used in IMRT and VMAT.[6] However, the lowering dose rate leads to prolonged treatment times, increasing the likelihood of patient motion during the treatment delivery[9] and necessitate more frequent patient monitoring with positional correction. The results of four factors, as shown in [Table 3], were not significantly different in terms of interplay effects affecting factors. These results were taken by an uncontrolled of an initial phantom moving phase during irradiation. This event engendered an unequal of number of counting detectors. Therefore, the gamma passing rate of these four factors was different. In other words, if the starting time of phantom moving during irradiation can be identically set, the percentage of gamma passing rate will be the same for four effect studies.

The results of clinical plans showed that the interplay effects have more influence on large dynamic motion, high plan complexity, and small field size. The complex plans groups provide worst results than the noncomplex plans groups. This point showed the similarity to a study from Netherton et al., whom found the maximum and mean dose deviations increased with increasing plan complexity under tumor motion.[24] According to [Table 4], clinical plan number one has a larger target volume than plan number two; however, the first patient was applied with a smaller field size (7.6 cm × 6.7 cm2) than the second patient (9.5 cm × 7.7 cm2). For this reason, the first clinical plan received a lower percentage of gamma passing rates. In addition, these clinical plans were created as the verification plans (QA plans). Thus, each plan was performed in only one fraction due to the limited practical time. However, we also anticipated that the number of fractions would be impacted to the interplay effect. This assumption was agreed with Stambaugh et al.[24] They revealed that the interplay effect was statistically significantly lower for the three-and five-fraction statistical simulations. Overall, the gradient effect dominates the clinical situation. Consequently, the interplay effect in clinical plans cannot be deliberated for some specific parameters but all dosimetric parameters of each case should be considered. Edvardsson et al. also indicated that the large interplay effect was observed for individual fraction and the extent varied with patient and machine-specific parameters. Each patient has a unique respiratory pattern that affects the particular planning parameters. These reasons make the different mutual movement between the target and the MLC, resulting in a different extent of the interplay effects.[9] For limitation of this study, because the 2D moving phantom applies with 2D diode array which can represent only planar doses. Thereby, this research cannot demonstrate an actual patient tumor motion and a tumor dose volume.


   Conclusions Top


The interplay effects for 6 MV FFF photon beams in lung cancer are more pronounced for the higher amplitudes and the smaller field sizes. The breathing amplitudes above 0.5 cm lead to significant changes in the shape of dose distribution due to the interplay effects, especially for 2D movements. Nevertheless, the interplay effects in real clinical cases cannot be cogitated with just a single or both parameters since the individual patient has specific parameters for target shape, target volume, number of targets, the distance between targets, planning parameters/techniques, and patient breathing pattern. Therefore, the entire dosimetric parameters of clinical cases have a critical impact on the outcomes of the interplay effects.

Acknowledgments

The authors would like to express our appreciation to all staff members of the Division of Radiation Oncology, King Chulalongkorn Memorial Hospital, Bangkok, Thailand, The CU graduate school thesis grant from Graduate School, Chulalongkorn University, Bangkok, Thailand and also thank Mr. Wasan Punyasang, M.Sc from Statistics, Research Affairs, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.

Financial support and sponsorship

The CU graduate school thesis grant from graduate school, Chulalongkorn University, Bangkok, Thailand.

Conflicts of interest

There are no conflicts of interest.

 
   References Top

1.
World Health Organization. Cancer. Geneva, Switzerland: World Health Organization; 2018. Available from: https://www.who.int/news-room/fact-sheets/detail/cancer. [Last accessed on 2020 Mar 01].  Back to cited text no. 1
    
2.
Sebastian NT, Xu-Welliver M, Williams TM. Stereotactic body radiation therapy (SBRT) for early stage non-small cell lung cancer (NSCLC): Contemporary insights and advances. J Thorac Dis 2018;10:S2451-64.  Back to cited text no. 2
    
3.
Tsang MW. Stereotactic body radiotherapy: Current strategies and future development. J Thorac Dis 2016;8:S517-27.  Back to cited text no. 3
    
4.
Janssen S, Käsmann L, Rudat V, Rades D. Stereotactic Body Radiation Therapy (SBRT) for recurrent non-small cell lung cancer (NSCLC). Anticancer Res 2016;36:825-8.  Back to cited text no. 4
    
5.
Mukhlisin M, Pawiro SA. Dosimetric impact of interplay effect in lung IMRT and VMAT treatment using in-house dynamic thorax phantom. J Phys Conf Ser 2016;694:012009.  Back to cited text no. 5
    
6.
Adamczyk M, Adamczyk S, Piotrowski T. Modelling the effects of lung cancer motion due to respiration. Nukleonika 2018;63:95-103.  Back to cited text no. 6
    
7.
Kubo K, Monzen H, Tamura M, Hirata M, Ishii K, Okada W, et al. Minimizing dose variation from the interplay effect in stereotactic radiation therapy using volumetric modulated arc therapy for lung cancer. J Appl Clin Med Phys 2018;19:121-7.  Back to cited text no. 7
    
8.
Kakakhel MB, Kairn T, Kenny J, Seet K, Fielding AL, Trapp JV. Interplay effects during enhanced dynamic wedge deliveries. Phys Med 2013;29:323-32.  Back to cited text no. 8
    
9.
Edvardsson A, Nordström F, Ceberg C, Ceberg S. Motion induced interplay effects for VMAT radiotherapy. Phys Med Biol 2018;63:085012.  Back to cited text no. 9
    
10.
Ehrbar S, Perrin R, Peroni M, Bernatowicz K, Parkel T, Pytko I, et al. Respiratory motion-management in stereotactic body radiation therapy for lung cancer – A dosimetric comparison in an anthropomorphic lung phantom (LuCa). Radiother Oncol 2016;121:328-34.  Back to cited text no. 10
    
11.
Riley C, Yang Y, Li T, Zhang Y, Heron DE, Huq MS. Dosimetric evaluation of the interplay effect in respiratory-gated RapidArc radiation therapy. Med Phys 2014;41:011715.  Back to cited text no. 11
    
12.
Vassiliev ON, Kry SF, Chang JY, Balter PA, Titt U, Mohan R. Stereotactic radiotherapy for lung cancer using a flattening filter free clinac. J Appl Clin Med Phys 2009;10:14-21.  Back to cited text no. 12
    
13.
Aoki S, Yamashita H, Haga A, Nawa K, Imae T, Takahashi W, et al. Flattening filter-free technique in volumetric modulated arc therapy for lung stereotactic body radiotherapy: A clinical comparison with the flattening filter technique. Oncol Lett 2018;15:3928-36.  Back to cited text no. 13
    
14.
Georg D, Knöös T, McClean B. Current status and future perspective of flattening filter free photon beams. Med Phys 2011;38:1280-93.  Back to cited text no. 14
    
15.
Sharma SD. Unflattened photon beams from the standard flattening filter free accelerators for radiotherapy: Advantages, limitations and challenges. J Med Phys 2011;36:123-5.  Back to cited text no. 15
[PUBMED]  [Full text]  
16.
Dang TM, Peters MJ, Hickey B, Semciw A. Efficacy of flattening-filter-free beam in stereotactic body radiation therapy planning and treatment: A systematic review with meta-analysis. J Med Imaging Radiat Oncol 2017;61:379-87.  Back to cited text no. 16
    
17.
Hazard LJ, Wang B, Skidmore TB, Chern SS, Salter BJ, Jensen RL, et al. Conformity of LINAC-based stereotactic radiosurgery using dynamic conformal arcs and micro-multileaf collimator. Int J Radiat Oncol Biol Phys 2009;73:562-70.  Back to cited text no. 17
    
18.
Miften M, Olch A, Mihailidis D, Moran J, Pawlicki T, Molineu A, et al. Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM Task Group No. 218. Med Phys 2018;45:e53-83.  Back to cited text no. 18
    
19.
Court LE, Wagar M, Ionascu D, Berbeco R, Chin L. Management of the interplay effect when using dynamic MLC sequences to treat moving targets. Med Phys 2008;35:1926-31.  Back to cited text no. 19
    
20.
Ong CL, Dahele M, Slotman BJ, Verbakel WF. Dosimetric impact of the interplay effect during stereotactic lung radiation therapy delivery using flattening filter-free beams and volumetric modulated arc therapy. Int J Radiat Oncol Biol Phys 2013;86:743-8.  Back to cited text no. 20
    
21.
Rao M, Ye J, Spiegel J, Takamiya R, Mehta VK, Wu J, et al. Interplay effect during lung SBRT with VMAT delivered by flattening filter-free beams. Int J Radiat Oncol Biol Phys 2014;90:S192.  Back to cited text no. 21
    
22.
Adamczyk M, Kruszyna-Mochalska M, Rucińska A, Piotrowski T. Software simulation of tumour motion dose effects during flattened and unflattened ITV-based VMAT lung SBRT. Rep Pract Oncol Radiother 2020;25:684-91.  Back to cited text no. 22
    
23.
Netherton T, Li Y, Nitsch P, Shaitelman S, Balter P, Gao S, et al. Interplay effect on a 6-MV flattening-filter-free linear accelerator with high dose rate and fast multi-leaf collimator motion treating breast and lung phantoms. Med Phys 2018;45:2369-76.  Back to cited text no. 23
    
24.
Stambaugh C, Nelms BE, Dilling T, Stevens C, Latifi K, Zhang G, et al. Experimentally studied dynamic dose interplay does not meaningfully affect target dose in VMAT SBRT lung treatments. Med Phys 2013;40:091710.  Back to cited text no. 24
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]



 

Top
Print this article  Email this article
  

    

 
   Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (1,220 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
   Subjects and Methods
   Results
   Discussion
   Conclusions
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed518    
    Printed16    
    Emailed0    
    PDF Downloaded113    
    Comments [Add]    

Recommend this journal