TECHNICAL NOTE
Year : 2014  |  Volume : 39  |  Issue : 4  |  Page : 247-250
 

Dose verification in carcinoma of uterine cervix patients undergoing 3D conformal radiotherapy with Farmer type ion chamber


1 Department of Radiotherapy and Oncology, Kasturba Medical College Hospital, Attavar, Mangalore, Karnataka, India
2 Medical Physics Unit, National Oncology Center, Royal Hospital, Muscat, Sultanate of Oman

Date of Submission05-Jun-2014
Date of Decision31-Jul-2014
Date of Acceptance01-Aug-2014
Date of Web Publication12-Nov-2014

Correspondence Address:
Challapalli Srinivas
Department of Radiotherapy and Oncology, Kasturba Medical College Hospital, Attavar, Mangalore - 575 001, Karnataka, India

Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0971-6203.144492

Rights and Permissions

 

   Abstract 

External beam radiotherapy (EBRT) for carcinoma of uterine cervix is a basic line of treatment with three dimensional conformal radiotherapy (3DCRT) in large number of patients. There is need for an established method for verification dosimetry. We tried to document absorbed doses in a group of carcinoma cervix patients by inserting a 0.6 cc Farmer type ion chamber in the vaginal cavity. A special long perspex sleeve cap is designed to cover the chamber for using in the patient's body. Response of ionization chamber is checked earlier in water phantom with and without cap. Treatment planning was carried out with X-ray computed tomography (CT) scan and with the chamber along with cap in inserted position, and with the images Xio treatment planning system. Three measurements on 3 days at 5-6 fraction intervals were recorded in 12 patients. Electrometer measured charges are converted to absorbed dose at the chamber center, in vivo. Our results show good agreement with planned dose within 3% against prescribed dose. This study, is a refinement over our previous studies with transmission dosimetry and chemicals in ampules. This preliminary work shows promise that this can be followed as a routine dose check with special relevance to new protocols in the treatment of carcinoma cervix with EBRT.


Keywords: 3D conformal radiotherapy, cancer cervix, dose verification; farmer chamber, in vivo dosimetry, ionization chamber


How to cite this article:
Srinivas C, Kumar SP, Ravichandran R, Banerjee S, Saxena P U, Arun Kumar E S, Pai DK. Dose verification in carcinoma of uterine cervix patients undergoing 3D conformal radiotherapy with Farmer type ion chamber. J Med Phys 2014;39:247-50

How to cite this URL:
Srinivas C, Kumar SP, Ravichandran R, Banerjee S, Saxena P U, Arun Kumar E S, Pai DK. Dose verification in carcinoma of uterine cervix patients undergoing 3D conformal radiotherapy with Farmer type ion chamber. J Med Phys [serial online] 2014 [cited 2023 Mar 26];39:247-50. Available from: https://www.jmp.org.in/text.asp?2014/39/4/247/144492



   Introduction Top


External beam radiotherapy (EBRT) is the most common standard treatment of choice in all non-metastatic stages of cancer of the uterine cervix. The ultimate check of the actual dose delivered to a patient in radiotherapy can only be achieved by using in vivo dosimetry [1] and this also serves as an important part of a quality assurance program that is recommended for improvement in quality of patient care. [2],[3] Though in vivo dosimetry is not routinely carried out in all patients on routine basis, it has application to detect or assess clinically relevant differences between planned and delivered dose and its potential to identify probable errors in dose calculation, data transfer, dose delivery, patient set-up. In special procedures such as stereotactic radiotherapy or intensity modulated radiotherapy, in-phantom dosimetry methods are used as pre-treatment verification. Semiconductor diodes and thermo luminescent dosimeters (TLD) are used in the principal techniques for in vivo dosimetry. [4],[5],[6],[7] Some other techniques of in vivo dosimetry employ metal oxide semiconductor field effect transistors, alanine, gels, plastic scintillators, radio chromic films, conventional portal films, or electronic portal imaging devices. [8],[9]

These detectors offer disadvantage owing to periodic calibrations, correction for temperature, pressure, photon energy dependence, accurate positioning, and estimation of the photon fluence perturbation inside the patient, and therefore used in a limited number of centers. The most direct form of in vivo dosimetry is to physically place detectors inside the patient. [10] When detectors are introduced in readily accessible body cavities, such as esophageal lumen, rectum, vagina and bladder, it is possible to measure the in vivo dose, [10],[11] and use of an insulated ion chamber for on line dose verification was first reported [12] in rectal cavity.

Earlier we reported chemical dosimeter (ferrous sulfate-benzoic acid-xylenol orange [FBX] aqueous chemical dosimeter system) ampules in vaginal cavity to record doses, and also used transmission measurements with ion chamber in a group of carcinoma cervix patients. [13],[14] We wanted to check the feasibility of using thimble chamber used for beam level dosimetry, for measuring absorbed doses in vivo, in the patients treated with three-dimensional conformal radiotherapy (3DCRT), as a dose verification method.


   Materials and Methods Top


Patients

A total number of 12 patients with carcinoma of uterine cervix were selected. An informed consent was taken from the patients about the in vivo dosimetric procedure by the physician. All the patients received treatment with linear accelerator with 3DCRT.

Detector used for dose verification

An absorbed dose calibrated Farmer type ionization chamber (IC) (FC65-G, IBA Dosimetry GmbH) was used to record doses in vivo. A specially designed perspex protection cap (total length 13cm and 1.5cm diameter) with an extended coverage till the end of aluminum stem of the IC [Figure 1] fabricated locally ensures better insulation during patient dosimetry by insertion into the vaginal cavity. Absolute dose measurements were conducted with IC in water phantom under source to axis distance (SAD) geometry, to find out response, depth dose dependency and field size variations, with and without the perspex cap.
Figure 1: Locally fabricated perspex protection cap with an extended coverage till the end of aluminum stem of the ionization chamber

Click here to view


Simulation and target volume delineation

All patients were immobilized with vacloc device (from M/s Klarity Medical, USA) in supine position with the hands placed overhead. During CT simulation procedure, ionization chamber with locally fabricated perspex cap covered with a thin sheath of rubber material (latex rubber) was inserted in the vagina (natural body cavity), maintaining all the necessary aseptic conditions for all patients, with head towards gantry. Scanned serial computed tomography (CT) images were exported to Focalsim contouring station (M/s Elekta Ltd., Crawly, UK) via digital imaging and communications in medicine (DICOM) network. The contouring of tumor volumes and normal structures was done by radiation oncologists. The clinical target volume (CTV) encompassed the gross tumor volume (GTV) with 5 mm margin together with pelvic nodal basins up to the bifurcation of inferior vena cava, was created. The planning target volume (PTV) was generated by adding 5 mm margin to the CTV to account for inter fractional geometric positional uncertainties. The region of IC was contoured in all transverse slices with an additional margin of 5 mm to account for probable inter fractional positional changes during the course of treatment.

Treatment planning and therapy execution

CMS XiO ® (Elekta Ltd, Crawly, UK) version 4.80.02 treatment planning system (TPS) utilizes Clarkson, convolution, superposition and fast superposition algorithms. The contoured images were transferred to TPS for beam placement and dose calculations by using superposition algorithm. A set of four beams with gantry angles 270 ° , 0 ° , 90 ° and 180 ° were placed and field in field technique was used as and when required to reduce the hot spots, and a 3DCRT plan was generated. A dose of 50 Gray in 25 fractions was prescribed to PTV which was normalized to the 100% isodose line. The mean dose calculated by TPS in the region of IC from the dose volume histogram (DVH) was noted. [Figure 2] represents the position of IC in the planning CT in transverse, coronal and sagittal planes with DVH. The 3DCRT plan was evaluated, finalized by the radiation oncologist and was exported to record and verification system (MOSAIQ ® ) for scheduling and execution. The lateral and anterior digitally reconstructed radiograph (DRR) images were exported to electronic portal imaging station (iViewC-camera based) for positional verification before treatment execution.
Figure 2: Isodose coverage around ionization chamber in the in transverse, coronal, and sagittal planes with dose volume histogram

Click here to view


All patients received 3DCRT with Elekta linac 6MV Photon energy, motorized wedge, 40 pairs of multi leaf collimator (MLCi2) with leaf thickness of 1cm projected at 100 cm isocenter and electronic portal imaging device (iViewC). Machine is calibrated for 1cGy/MU at iso center at a dose rate of 350 MU/min.

Prior to treatment execution, verification of patient's treatment setup under LINAC was checked with iViewC. A 3mm margin of translational (x, y, and z) errors were permitted and necessary couch shifts were applied, as and when required. The scheduled 3DCRT treatment was subsequently executed. [Figure 3]a and b show the portal images of anterior and lateral treatment fields.
Figure 3: (a) Anterior treatment portal image. (b) Lateral treatment portal image

Click here to view


Dose measurements

Treated patients had inserted IC in vaginal cavity as described during simulation. Temperature of patient's body is taken at the time of treatment. Electrometer (Dose1, Iba) readings were recorded in the proforma (in nano coulombs, nC) during treatment execution. With necessary correction factors (N D,w , temperature of body, pressure, polarity, beam quality & saturation) the absorbed dose (in Gy) at the dosimeter locations are recorded. Three measurements were taken for each patient with scheduled gap of 5-6 fractions intervals. Mean dose of all three measurements were taken to compare with the planned dose at the reference point of the detector.


   Results Top


The median age of studied patients was 58 years. Of the 12 subjects, five had stage IIB and seven had stage IIIB disease. The measured doses with IC in water phantom in SAD geometry, field size and depth dose dependency with and without prototype perspex cap is shown in [Table 1]. It was observed that there was no significant difference in the absolute doses with water medium around, with or without perspex cap used for patient's measurements. Corrected measurements at 10 cm depth also were in good agreement with TPS dose values used for patient's treatment plans using tissue maximum ratios. [Table 2] shows the estimates of absorbed doses in vivo, for all the 12 patients. Mean doses with IC recorded in patients were in good agreement with planned doses in these patients by treatment planning system (Mean DVH at reference point of measurement). The percentage deviation of measured dose (with IC) with calculated dose by TPS (from DVH) of individual patients ranges from -3.05 to 2.96%. [Table 3] showed the Comparison of results (mean % deviation with standard deviation for measured vs calculated dose) against our previous reports. [13],[14]
Table 1: Absolute dose (Gy) measured measured with IC for different field sizes and depths with and without perspex cap using variable depth water phantom under SAD condition


Click here to view
Table 2: The % deviation of measured dose (with IC) with calculated dose by TPS (from DVH)


Click here to view
Table 3: Comparison of in vivo dose estimates with previous reports (mean % deviation with SD for measured vs calculated dose)


Click here to view



   Discussion Top


An effective way of checking the quality of the entire dosimetric procedure, from the accurate positioning of patient to the performance of the treatment machine, is to take absorbed dose measurements on the patient and, when possible, in natural body cavities. [15] Our present work has brought out an effective way of documentation of absorbed dose in clinical radiotherapy, which has application where such cavities exist. Also, 0.6 cc Farmer type ion chambers are available in all medical physics departments, and therefore any verification methods could be designed with this detector. As field sizes used in pelvic treatments are broad enough to give an accurate estimate of central doses, this method could be adopted in intensity modulated treatments for prostate cancers where placement of dosimeter in rectum is feasible. Entrance and exit dose measurements with diode or TLDs are common in total body irradiation with photons and such measurements have been done prostate, bladder and parotid gland tumor treatments. [Table 3] highlights that our present work has solved the problem in reducing the deviation encountered in our previous studies, [13],[14] because during that period of time, treatments were based on simple plans with measured inter-field separations and there were no documented methods for DVH averaging at the position of detectors. As 0.6 cc chambers average doses across its air volume, smaller IC detectors could be employed if segmental small fields are encountered in treatment execution. As the IC is positioned at the measuring position at the time of simulation, taken into account for pixel based treatment planning, the reproducibility is well ensured, and therefore we confirm that our measured doses are fairly accurate enough for documentation.


   Conclusions Top


Dose verification before starting and during the course is becoming essential part of modern radiotherapy. Present study evaluated the applicability of a conventional ionization chamber for performing in vivo dose measurements in patients of carcinoma of uterine cervix undergoing 3DCRT. As a verification procedure this could be incorporated in clinical trials if it is warranted. Future work is needed to compare this simple method with well established small field dosimeters like diamond detectors.


   Acknowledgment Top


The authors wish to thank and gratefully acknowledge the comments/suggestions offered during the review of this paper by Dr. S.D. Sharma, RPAD, Mumbai, Dr. S. Sathiyan, KMIO, Bangalore & Mr. Sowmyanarayanan, Videhi Institute of Medical Sciences, Bangalore.

 
   References Top

1.
ICRU Report 24. Determination of dose in patient irradiated by means of X or gamma rays in radiotherapy procedures. International Commission on Radiation Units and Measurements. Bethesda; 1976.  Back to cited text no. 1
    
2.
Ferreira IH, Dutreix A, Bridier A, Chavaudra J, Svensson H. The ESTRO-Quality assurance network (EQUAL). Radiother Oncol 2000;55:273-84.  Back to cited text no. 2
    
3.
The Royal College of Radiologists, Society and College of Radiographers, Institute of Physics and Engineering in Medicine, National Patient Safety Agency, British Institute of Radiology. Towards Safer Radiotherapy. London: The Royal College of Radiologists; 2008.  Back to cited text no. 3
    
4.
Huyskens DP, Bogaerts R, Verstraete J, Loof M, Nystrom H, Fiorino C, et al. Practical guidelines for the implementation of in vivo dosimetry with diodes in external radiotherapy with photon beams (entrance dose). ESTRO. Brussel; 2001.  Back to cited text no. 4
    
5.
van Dam J, Marinello G. Methods for in vivo dosimetry in external radiotherapy, ESTRO, Brussel; 1994.  Back to cited text no. 5
    
6.
Kron T. Applications of thermo luminescence dosimetry in medicine. Radiat Prot Dosim 1999;85:333-40.  Back to cited text no. 6
    
7.
Mayles WP, Heisig S, Mayles HM. Treatment verification and in vivo dosimetry. In: Williams JR, Thwaites DI, editors. Radiotherapy physics: In practice. Oxford Univ.Press; 2000. p. 220-46.  Back to cited text no. 7
    
8.
Evans P, Marinello G. Quality control of treatment delivery. In: Mayles P, Nahun A, Rosenwald JC, editors. Handbook of radiotherapy physics: Theory and practice. New York: Taylor and Francis; 2007. p. 867-95.  Back to cited text no. 8
    
9.
In vivo dosimetry using MOSFET detectors in radiotherapy. In: Gurp EB, editor. Universitaire Pers Maastricht; 2009.  Back to cited text no. 9
    
10.
Podgorsak EB. Radiation Oncology Physics: A Handbook for Teachers and Students, International Atomic Energy Agency, Vienna; 2005.  Back to cited text no. 10
    
11.
van Dam J, Marinello G. Methods for In Vivo dosimetry in External Radiotherapy. Italy: ESTRO Publication; 2006.  Back to cited text no. 11
    
12.
Marcie S, Charpiot E, Bensadoun RJ, Ciais G, Herault J, Costa A, et al. In vivo measurements with MOSFET detectors in oropharynx and nasopharynx intensity-modulated radiation therapy. Int J Radiat Oncol Biol Phys 2005;61:1603-6.  Back to cited text no. 12
    
13.
Srinivas C, Shenoy KK, Dinesh M, Savitha KS, Kasturi DP, Supe S, et al. In-vivo dosimetric study of carcinoma of uterine cervix with FBX solution in external beam therapy. J Med Phys 1999;24:47-50.  Back to cited text no. 13
  Medknow Journal  
14.
Srinivas C, Shenoy KK, Shetty J, Solomon JGR, Ravichandran R, Supe S, et al. In-vivo dosimetry: Confirmation of dose delivery through transit dose measurements in carcinoma of uterine cervix patients undergoing teletherapy. J Med Phys 2004;29:177-78.  Back to cited text no. 14
    
15.
World Health Organization, Quality Assurance in Radiotherapy, WHO, Geneva; 1988.  Back to cited text no. 15
    


    Figures

  [Figure 1], [Figure 2], [Figure 3]
 
 
    Tables

  [Table 1], [Table 2], [Table 3]


This article has been cited by
1 Simultaneous measurement of In vivo and transit mid-plane doses with ionization chambers in gynecological malignancy patients undergoing three-dimensional conformal radiotherapy
PuthaSuman Kumar, Challapalli Srinivas, BM Vadhiraja, Sourjya Banerjee, R Shreyas, PU Prakash Saxena, Ramamoorthy Ravichandran, Dilson Lobo
Journal of Medical Physics. 2020; 45(2): 123
[Pubmed] | [DOI]



 

Top
Print this article  Email this article
  

    

 
   Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (691 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusions
   Acknowledgment
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed2475    
    Printed77    
    Emailed1    
    PDF Downloaded256    
    Comments [Add]    
    Cited by others 1    

Recommend this journal