Influence of increment of gantry angle and number of arcs on esophageal volumetric modulated arc therapy planning in Monaco planning system: A planning study


1 Department of Radiation Oncology, Dharamshila Hospital and Research Centre, New Delhi, India
2 School of Advanced Sciences, Vellore Institute of Technology University, Vellore, Tamil Nadu, India

Date of Submission08-May-2014
Date of Decision06-Sep-2014
Date of Acceptance06-Sep-2014
Date of Web Publication12-Nov-2014

Correspondence Address:
L Nithya
Chief Medical Physicist and RSO, Dharamshila Hospital and Research Centre, Vasundhara Enclave, New Delhi 110 096
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0971-6203.144488

Rights and Permissions

 

   Abstract 

The objective of this study was to analyze the influence of the increment of gantry angle and the number of arcs on esophageal volumetric modulated arc therapy plan. All plans were done in Monaco planning system for Elekta Synergy linear accelerator with 80 multileaf collimator (MLC). Volumetric modulated arc therapy (VMAT) plans were done with different increment of gantry angle like 15 o , 20 o , 30 o and 40 o . The remaining parameters were similar for all the plans. The results were compared. To compare the plan quality with number of arcs, VMAT plans were done with single and dual arc with increment of gantry angle of 20 o . The dose to gross tumor volume (GTV) for 60 Gy and planning target volume (PTV) for 48 Gy was compared. The dosimetric parameters D 98% , D 95% , D 50% and D max of GTV were analyzed. The homogeneity index (HI) and conformity index (CI) of GTV were studied and the dose to 98% and 95% of PTV was analyzed. Maximum dose to spinal cord and planning risk volume of cord (PRV cord) was compared. The Volume of lung receiving 10 Gy, 20 Gy and mean dose was analyzed. The volume of heart receiving 30 Gy and 45 Gy was compared. The volume of normal tissue receiving greater than 2 Gy and 5 Gy was compared. The number of monitor units (MU) required to deliver the plans were compared. The plan with larger increment of gantry angle proved to be superior to smaller increment of gantry angle plans in terms of dose coverage, HI, CI and normal tissue sparing. The number of arcs did not make any difference in the quality of the plan.


Keywords: Esophagus cases; increment of gantry angle; number of arcs; VMAT


How to cite this article:
Nithya L, Nambi Raj N A, Rathinamuthu S, Sharma K, Pandey MB. Influence of increment of gantry angle and number of arcs on esophageal volumetric modulated arc therapy planning in Monaco planning system: A planning study. J Med Phys 2014;39:231-7

How to cite this URL:
Nithya L, Nambi Raj N A, Rathinamuthu S, Sharma K, Pandey MB. Influence of increment of gantry angle and number of arcs on esophageal volumetric modulated arc therapy planning in Monaco planning system: A planning study. J Med Phys [serial online] 2014 [cited 2023 Mar 26];39:231-7. Available from: https://www.jmp.org.in/text.asp?2014/39/4/231/144488



   Introduction Top


Volumetric Modulated Arc Therapy (VMAT) is a dynamic arc delivery technique of intensity modulated radiation therapy (IMRT). In VMAT, gantry speed, multileaf collimator (MLC) shape and speed, dose rate and collimator are continuously changing according to the treatment plan. VMAT has the potential benefits compared to IMRT in terms of tumor control probability (TCP) and reducing the toxicity to the normal structures [1],[2] with high monitor unit (MU) and treatment delivery efficiency. [1],[2],[3],[4] The quality of VMAT plan depends on the algorithm used for optimization and dose calculation, the physical and biological parameters used in the treatment planning system (TPS). Many authors have studied the influence of different physics and machine parameters on the VMAT plan quality using Pinnacle and Eclipse planning system. [4],[5] We carried out our study on Monaco planning system to analyze the influence of Some of the physical parameters on VMAT plan quality.

Monaco planning system supports IMRT as well as VMAT planning. It uses biological models for optimization. [6],[7] Monaco uses machine parameters like leaf speed, gantry speed and dose rate for optimization. One of the parameters is the Increment of gantry angle (IGA). IGA controls the number of generated static gantry positions or sectors in VMAT plan. One more parameter is number of arcs. The number of sectors or the number of arcs used for planning may affect the quality of the plan. Yin L et al., [1] and Abbas et al., [8] compared the quality of the esophageal VMAT plans with number of arcs using Eclipse and Pinnacle planning systems respectively. Yang et al., [4] compared single arc and double arc VMAT plans for head and neck, prostate, lung and spine using Pinnacle planning system.

Twelve middle third esophageal patients were taken for this study. All these cases were planned in Monaco planning system for different IGA such as 15 o , 20 o , 30 o and 40 o which are the angles we use for daily routine planning. Also separate plans were done for single and dual arc with IGA 20 o which is the angle mostly used for treatment planning. Tumor dose and dose to normal structures were analyzed. In this study, the influence of IGA and number of arcs in quality of VMAT plan for esophageal cases was analyzed.


   Materials and Methods Top


Patient selection

A total of 12 cases comprising middle third esophageal cancer patients were considered for this study. The median age was 59 (Range = 40-75 years) with 9 males and 3 females. The clinical stage distribution was T3 in 10 patients, and T4 in 2 patients with node stage N1-N3. Histopathology was squamous cell carcinoma for all patients. The mean length of GTV was 9.9 cm (Range = 6.3-12.3 cm) and PTV was 16 cm (Range = 12.5-20.5 cm) with mean volume of 95 cc (Range = 40-177) and 584 cc (Range = 310-718 cc) respectively.

Monaco planning system

Monaco treatment planning system (Elekta Ltd, Crawly, UK) version 3.20.02 utilizes physical effects of radiation and biological properties of the tissue. It has three biological constraints such as Target EUD, Parallel and serial and six physical constraints such as target penalty, quadratic overdose, overdose DVH, under dose DVH, maximum dose and quadratic under dose. The user has an option to set the cell sensitivity of the tumor in target EUD. The organ at risk can be set as serial or parallel constraints depending on the properties of the tissue.

The system uses a two-stage process of optimizing dose distribution. Generally, in stage one the ideal fluence distribution of beams is optimized to meet a user-defined prescription for given set of beams. In stage two, segmentation is done, which includes the segment shapes and weights, so that deliverable fields are obtained. In this stage system uses Monte Carlo simulation during optimization.

In VMAT optimization, prior to stage one, system divides the beam into sectors. In stage one, at the initialization stage, the system creates the dose calculation cube around all defined structures and calculates structure volumes using cubic voxels. Then it projects the union of all target volumes with the margin defined. Numbers of static sectors are created based on arc length and user defined IGA. Beamlets for each sector are created. Width of beamlet is user defined and length is equal to the length of individual MLC leaves. The system uses an enhanced pencil beam algorithm to calculate the open field dose. Then, the fluence optimization begins in which the weights (fluence) of all individual pencil beams are varied simultaneously. The unconstrained problems are solved by conjugate gradient algorithm. After the unconstrained optimization finishes, if necessary the system changes each cost function relative weight to make the optimizer meet the isoconstraints and restarts the unconstrained optimization problem. Stage one optimization continues until all the constraints are met. The accuracy of dose at the end of the stage one is limited because the algorithm is kernel based two dimensional method, especially in the presence of heterogeneities.

In second stage, the treatment planning system considers the deliverability of accelerator. It takes each fluence map and sequences it in such a way that it is spread over the original sector it represents. The system determines leaf trajectories based on the target dose rate defined by the user. If segment shape optimization (SSO) method is selected, the system selects the optimal dose rate by its own. Then, the system converts optimized fluences into deliverable arc sequence with multiple control points and the gantry position. The gantry positions need not be equally spaced. Dose calculation is done with voxel based Monte Carlo algorithm. The user can change the calculation accuracy and time by modifying some parameters like dose rate, Monte Carlo grid spacing and variance.

Increment of gantry angle and sector

IGA controls the number of generated static gantry positions or sectors. The user has to determine the number of sectors by dividing the arc length by increment. For example a 360 o arc length with 30 o IGA equals 12 static sectors to optimize as shown in [Figure 1]. Prior to stage one optimization, the system divides a sequence into sectors that are used for simulating the arc during stage one optimization. The system generates fluence maps during stage one and computes them at IGA. Generally, using a large IGA creates few sectors which can produce poor quality plans and increase treatment time and using a too small IGA gives more sectors which may increase the quality of the plan.
Figure 1: Sectors for 360° arc length with 12 sectors

Click here to view


The number of sectors plays a role in the leaf movement. Monaco treatment planning system uses sweep sequencer for VMAT. During first stage optimization, the sequencer reorders the fluence profiles along with sectors. The leaf movement direction alternates between sectors that is the leaves at the left field edge in one sector move to the right field edge as the gantry rotates. The leaf edges arrive at the field edge at the beginning of the next sector where they change the direction.

Imaging and contouring

All patients were immobilized with thermoplastic sheet in supine position and the hands were kept above the head. The computed tomography (CT) images were acquired for all patients in Biograph PET-CT (Siemens AG, Medical solutions, Germany) with 3 mm slice thickness; field of view 50 cm. CT was acquired in normal breathing position. No gating or respiratory assistant instruments were used. The images were imported into Monaco treatment planning system via DICOM and the contouring of tumor volumes and normal structures were done by the radiation oncologists.

Target volumes

The gross tumor volume encompassed the esophageal tumor and the lymph node was included if it was positive. The regional lymph nodes were drawn as clinical target volume (CTV). The CTV was expanded with 8 mm in all six directions to create planning target volume (PTV) to manage the setup uncertainties. The dose prescription for GTV was 60 Gy and PTV was 48 Gy.

Normal structures

The normal structure comprised of both right and left lung, heart and spinal cord. The spinal cord was expanded by 5 mm to create PRV cord. The body contour minus all tumor volumes and normal structures was taken as normal tissue.

Dose prescription and acceptance parameters

All plans were generated to deliver 60 Gy (2 Gy per fraction) to GTV and 48 Gy (1.6 Gy per fraction) to PTV in 30 fractions. The primary goal of treatment planning was to cover 95% volume of GTV and PTV with the prescribed dose of 60 Gy and 48 Gy respectively and to restrict not more than 10% volume of GTV to receive 107% of prescribed dose (64.2 Gy). The maximum dose to spinal cord and PRV cord was restricted to 45 Gy and 50 Gy respectively. The mean dose to total lung should be less than 20 Gy. The 30% of total lung should not receive more than 20 Gy (V 20 ≤ 30%) and 60% should not receive more than 10 Gy (V 10 ≤ 60%). The 33% of heart should not receive more than 45 Gy (V 45 ≤ 33%) and 67% should not receive more than 30 Gy (V 30 ≤ 67%). All planning parameters are summarized in [Table 1].
Table 1: Treatment planning objectives


Click here to view


Planning parameters used in TPS

All VMAT plans were planned with the following calculation properties: Grid spacing was selected as 3 mm, and Monte Carlo variance was 3%. Monte Carlo algorithm was selected as secondary algorithm for second stage dose calculation that is final dose calculation. The dose was calculated to the medium and not to the water. For all plans, heterogeneity correction was applied.

All plans were planned for 360 o gantry rotation with single arc in clock wise direction from 180 o -180 o . The IGA was taken as 15 o , 20 o , 30 o and 40 o and the plans were named as VMAT 15 , VMAT 20 , VMAT 30 and VMAT 40 respectively. Remaining parameters kept similar. Segment shape optimization method was used. Minimum segment width was kept as 5 mm and the fluence smoothing level was at medium level.

In Monaco planning system, multiple arcs can be created. We can create multiple arcs with different beam parameters or dual arc for same beam parameters. In this study all plans were planned with single arc (VMAT S ) with clockwise direction and dual arc (VMAT D ) for the same beam with clockwise and counter clockwise directions.

Plan comparison

Evaluation parameters

The comparison of all VMAT plans was evaluated using the following terms:

  • Homogeneity Index (HI): ( D 2% - D 98% )/D 50% , a ratio evaluating the dose homogeneity in GTV where D 2% , D 98% and D 50% are the minimum dose delivered to 2%, 98% and 50% volume of the GTV respectively. [9] HI of zero indicates the dose distribution is homogeneous.
  • Conformity Index (CI) : V pres /TV p , a ratio evaluating the coverage of the prescription dose in treatment plan, where V pres was the volume of body receiving the prescribed dose and TV p was the volume of GTV receiving the prescription dose [10] CI of one indicates the good dose conformity.
  • Target volumes: D 98% , D 95% , D 50% and D max for GTV were analyzed, where D 98% , D 95% , D 50% is minimum dose delivered to 98%, 95% and 50% Volume of GTV respectively and D max is the maximum dose.
  • Normal structures: D max for spinal cord and PRV cord were analyzed, where D max is the maximum dose. Both lungs were analyzed for V 20Gy , V 30Gy and mean lung dose. The heart dose was analyzed for V 30Gy and V 45Gy .V x represents the volume of organ irradiated above the x dose.
  • Normal tissue dose: The volume of normal tissue receiving ≥2 Gy and ≥5Gy were analyzed.
  • Number of monitor units: The number of monitor units required to deliver the plan was analyzed for all plans.


Statistical analyses

Statistical tests of significance were used for calculating the differences between VMAT plans. Wilcoxon matched-pair signed-rank test (two tailed, P ≤ 0.05) was used.


   Results Top


Increment of gantry angle


In this study, we examined 12 middle third esophageal cases using VMAT plans with different IGA such as 15 o , 20 o , 30 o and 40 o . All plans met our dosimetric criteria. We evaluated the plans using dose volume histogram (DVH). However we observed some differences between the plans. The dosimetric comparisons of these plans are summarized. [Table 2] shows the average value and standard deviation for all the 12 cases. Isofill distribution of one of the patients was given in [Figure 2]. DVH comparison of VMAT plans with different IGA was shown in [Figure 3].
Figure 2: Isofill comparison of VMAT plans with increment of gantry angle in axial, coronal and sagital view

Click here to view
Figure 3: DVH comparison of VMAT plans with different increment of gantry angle

Click here to view
Table 2: Statistical comparison of dosimetric parameters for (N=12) plans with different increment of gantry angle


Click here to view


There was no statistical significance differences were observed between VMAT 15 , VMAT 20 , VMAT 30 and VMAT 40 plans in dosimetric parameter of GTV such as D 98% , D 95% and D 50%. VMAT 15 had a higher D max and V 63 while comparing other plans. VMAT 30 had superior HI with good conformity. VMAT 15 plan was more heterogeneous plan than other plans. PTV coverage was good in all plans. Dosimetric parameters were similar in VMAT 30 and VMAT 40 and no statistical significant difference were observed.

Dose to total lung was analyzed. V 20Gy was in favour of VMAT 40 and remaining plans were similar. V 10Gy was high in VMAT 30 and mean lung dose was similar in all plans and did not show any statistical significant differences between them. Dose to spinal cord met the dosimetric criteria and no statistical significant difference was observed. Dose to PRV cord was in favor of VMAT 30 .

We analyzed the volume of normal tissue receiving doses ≥2 Gy and ≥5 Gy between plans. The volume of normal tissue receiving ≥2 Gy was almost same in all plans, however the volume receiving ≥5 Gy was high in higher IGA that is VMAT30 and VMAT 40.

The number of MUs required to deliver the plan was reduced by 20%-30% in VMAT 30 and VMAT 40 plans than VMAT 15 and VMAT 20.

Number of Arcs

In this study, we also analyzed all 12 cases using single vs. double arc VMAT plans. Dose to GTV showed no statistical significant difference in dosimetric parameters like D 98% , D 95% , D 50% , V 63 , HI and CI. Dose to PTV was similar in all dosimetry aspects. Dose to total lung, heart, spinal cord and PRV cord were similar and no statistical significant difference were observed. While looking into normal tissue dose single arc plan was superior to double arc plan. The results are summarized in [Table 3]. Isofill comparison of VMAT plan with single arc and double arc was shown in [Figure 4] and DVH comparison was shown in [Figure 5].
Figure 4: Isofill comparison of plans with single arc and double arc in axial, coronal and sagital view

Click here to view
Figure 5: DVH comparison of VMAT plans with single and double arc

Click here to view
Table 3: Statistical comparison of dosimetric parameters for (N=12) plans with single and double arc


Click here to view


The number of MUs required to deliver the plan in single arc was only 7% greater than that of double arc plan. We did not observe any statistical significant difference in MUs.


   Discussion Top


In this study, we selected 12 middle third esophageal cases which were already treated by VMAT or IMRT plans. The plans were done with different IGA keeping other parameters were same. The dose distribution and dosimetric parameters were acceptable according to our goal. General concept is, the smaller IGA gives better plan than larger IGA. In our study, we had observed that both smaller and larger IGA were giving good coverage and acceptable plans. D max and V 63 were high in VMAT 15 due to finer modulation compared to other plans. VMAT 30 gave very homogeneous plan but increased the dose to lung. Also we observed that finer modulation increased the number of MUs by 20-30%.

Abbas et al., had done the comparative study on single vs. double arc with Pinnacle planning system. They reported that there was no difference found in dosimetric parameters. They got more MUs for double arc plan (332 MU) than single arc plan (309 MU). In our study, we could not find any differences in dosimetric parameters. However, We differ in MU point of view, that is we got lesser MUs for double arc (606) than single arc (653).


   Conclusion Top


Monaco treatment planning system was capable of generating efficient VMAT plans. The quality of VMAT plan for all four increment of gantry angle was comparable. With good dose coverage, homogeneity index and increased MU efficiency we can use larger increment of gantry angle for middle third esophageal cases but finding may change the studies on other sites and would require more research on this concept.

 
   References Top

1.
Yin L, Wu H, Gong J, Geng JH, Jiang F, Shi AH, et al. Volumetric-modulated arc therapy vs. c-IMRT in esophageal cancer: A treatment planning comparison. World J Gastroenterol 2012;18:5266-75.  Back to cited text no. 1
    
2.
Hawkins MA, Bedford JL, Warrington AP, Tait DM. Volumetric modulated arc therapy planning for distal oesophageal malignancies. Br J Radiol 2012;85:44-52.  Back to cited text no. 2
    
3.
Van Benthuysen L, Hales L, Podgorsak MB. Volumetric modulated arc therapy vs. IMRT for the treatment of distal esophageal cancer. Med Dosim 2011;36:404-9.  Back to cited text no. 3
    
4.
Yang K, Yan D, Tyagi N. Sensitivity analysis of physics and planning SmartArc parameters for single and partial arc VMAT planning. J Appl Clin Med Phys 2012;13:3760.  Back to cited text no. 4
    
5.
Mihaylov IB, Curran B, Sternick E. The effect of gantry spacing resolution on plan quality in a single modulated arc optimization. J Appl Clin Med Phys 2011;12:3603.  Back to cited text no. 5
    
6.
Impac Medical Systems Inc. Monaco training guide. Sweden: Elekta AB, Stockholm; 2013.  Back to cited text no. 6
    
7.
Allen Li X, Alber M, Deasy JO, Jackson A, Ken Jee KW, Marks LB, et al. The use and QA of biologically related models for treatment planning: Short report of the TG-166 of the therapy physics committee of the AAPM. Med Phys 2012;39:1386-409.  Back to cited text no. 7
    
8.
Abbas AS, Moseley D, Kassam Z, Kim SM, Cho C. Volumetric-modulated arc therapy for the treatment of a large planning target volume in thoracic esophageal cancer. J Appl Clin Med Phys 2013;14:4269.  Back to cited text no. 8
    
9.
ICRU Report 83. Prescribing, recording, and reporting photon-beam intensity modulated radiation Therapy (IMRT). International Commission on Radiation Units and Measurements, Bethestha. 2010  Back to cited text no. 9
    
10.
Pathak P, Vashisht S. A quantitative analysis of intensity-modulated radiation therapy plans and comparison of homogeneity indices for the treatment of gynecological cancers. J Med Phys 2013;38:67-73.  Back to cited text no. 10
[PUBMED]  Medknow Journal  


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5]
 
 
    Tables

  [Table 1], [Table 2], [Table 3]


This article has been cited by
1 Dosimetric characteristics of VMAT plans with respect to a different increment of gantry angle size for Ca cervix
Munirathinam Natraj, P. N. Pawaskar, Arun Chairmadurai
Journal of Radiotherapy in Practice. 2020; : 1
[Pubmed] | [DOI]
2 The influence of increment of gantry on VMAT plan quality for cervical cancer
Along Chen, Zhenghuan Li, Li Chen, Maosheng Lin, Bin Li, Fei Chen
Journal of Radiation Research and Applied Sciences. 2019; 12(1): 447
[Pubmed] | [DOI]



 

Top
Print this article  Email this article
  

    

 
   Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (1,782 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusion