ORIGINAL ARTICLE
Year : 2008  |  Volume : 33  |  Issue : 3  |  Page : 108-113

Absolute dose determination in high-energy electron beams: Comparison of IAEA dosimetry protocols


Department of Radiation Physics, Kidwai Memorial Institute of Oncology, Bangalore, India

Correspondence Address:
S Sathiyan
Department of Radiation Physics, Kidwai Memorial Institute of Oncology, Bangalore - 560 029
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0971-6203.42754

Rights and Permissions

In this study, absorbed doses were measured and compared for high-energy electrons (6, 9, 12, 16, and 20 MeV) using International Atomic Energy Agency (IAEA), Technical Reports Series No. 277 (TRS), TRS 381, and TRS 398 dosimetry protocols. Absolute dose measurements were carried out using FC65-G Farmer chamber and Nordic Association of Clinical Physicists (NACP) parallel plate chamber with DOSE1 electrometer in WP1-D water phantom for reference field size of 15 x 15 cm 2 at 100 cm source-to-surface distance. The results show that the difference between TRS 398 and TRS 381 was about 0.24% to 1.3% depending upon the energy, and the maximum difference between TRS 398 and TRS 277 was 1.5%. The use of cylindrical chamber in electron beam gives the maximum dose difference between the TRS 398 and TRS 277 in the order of 1.4% for energies above 10 MeV (R 50 > 4 g/cm 2 ). It was observed that the accuracy of dose estimation was better with the protocols based on the water calibration procedures, as no conversion quantities are involved for conversion of dose from air to water. The cross-calibration procedure of parallel plate chamber with high-energy electron beams is recommended as it avoids p wall correction factor entering into the determination of k Q,Qo .


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed7105    
    Printed242    
    Emailed3    
    PDF Downloaded722    
    Comments [Add]    
    Cited by others 2    

Recommend this journal