|
ORIGINAL ARTICLE |
|
|
|
Year : 2006 | Volume
: 31
| Issue : 2 | Page : 78-82 |
|
Evaluation of gafchromic EBT film for intensity modulated radiation therapy dose distribution verification
A Sankar1, PG Goplakrishna Kurup2, V Murali2, Komanduri M Ayyangar3, R Mothilal Nehru3, J Velmurugan4
1 Department of Radiotherapy, Apollo Speciality Hospital, Chennai; Department of Physics, Anna University, Chennai, India 2 Department of Radiotherapy, Apollo Speciality Hospital, Chennai, India 3 Department of Radiotherapy, University of Nebraska Medical Center, Omaha, US 4 Department of Physics, Anna University, Chennai, India
Correspondence Address: A Sankar Department of Radiotherapy, Apollo Specialty Hospital, Padma Complex, 320 Mount Road, Chennai - 600 025, India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/0971-6203.26693
Abstract | | |
This work was undertaken with the intention of investigating the possibility of clinical use of commercially available self-developing radiochromic film - Gafchromic EBT film - for IMRT dose verification. The dose response curves were generated for the films using VXR-16 film scanner. The results obtained with EBT films were compared with the results of Kodak EDR2 films. It was found that the EBT film has a linear response between the dose ranges of 0 and 600 cGy. The dose-related characteristics of the EBT film, like post-irradiation color growth with time, film uniformity and effect of scanning orientation, were studied. There is up to 8.6% increase in the color density between 2 and 40 h after irradiation. There was a considerable variation, up to 8.5%, in the film uniformity over its sensitive region. The quantitative difference between calculated and measured dose distributions was analyzed using Gamma index with the tolerance of 3% dose difference and 3 mm distance agreement. EDR2 films showed good and consistent results with the calculated dose distribution, whereas the results obtained using EBT were inconsistent. The variation in the film uniformity limits the use of EBT film for conventional large field IMRT verification. For IMRT of smaller field size (4.5 x 4.5 cm), the results obtained with EBT were comparable with results of EDR2 films.
Keywords: EDR2 film, intensity modulated radiotherapy, radiochromic film
How to cite this article: Sankar A, Goplakrishna Kurup P G, Murali V, Ayyangar KM, Mothilal Nehru R, Velmurugan J. Evaluation of gafchromic EBT film for intensity modulated radiation therapy dose distribution verification. J Med Phys 2006;31:78-82 |
How to cite this URL: Sankar A, Goplakrishna Kurup P G, Murali V, Ayyangar KM, Mothilal Nehru R, Velmurugan J. Evaluation of gafchromic EBT film for intensity modulated radiation therapy dose distribution verification. J Med Phys [serial online] 2006 [cited 2023 Mar 29];31:78-82. Available from: https://www.jmp.org.in/text.asp?2006/31/2/78/26693 |
The quantitative dose validation of intensity modulated radiotherapy (IMRT) requires the three-dimensional high-resolution dosimetry systems with uniform sensitivity over its entire sensitive region. Silver halide radiographic films offer good resolution for two-dimensional dosimetry. Three-dimensional verification can be accomplished by measuring at different planes using films. The spectral sensitivity and response to the developing conditions are the important limiting factors of the silver halide films.[1] Self-developing radiochromic films offer high-resolution dose measurements with relatively insensitive spectral response.[2] Since these films are self-developing, the uncertainties associated with the developing conditions are ruled out. The present study deals with the evaluation of new type of radiochromic films - Gafchromic external beam therapy (EBT) film - for its use of IMRT plan validation. The results obtained with EBT films are compared with the results of commonly used Kodak extended dose range 2 (EDR2) films.
Materials and Methods | |  |
Film calibration
The calibration data set for both EDR2 and EBT films were obtained by exposing the films with doses ranging from 19.6 cGy to 700 cGy using 6MV photon beams. The film was kept at a dmax in solid water phantom in a plane perpendicular to the beam axis. The processed EDR2 films were scanned using VIDAR 16 (VXR-16) digitizer and RIT 113 film dosimetry software. The irradiated EBT films were scanned 3 h after irradiation along portrait direction to allow the saturation of color growth, as recommended by the film manufacturer. The EBT films are scanned with the yellow filter supplied by the film manufacturer to enhance the sensitivity of the film with VXR 16 scanner as the EBT film strongly absorbs the higher wavelength region of the visible spectrum. The measured optical density (OD) was plotted as a function dose [Figure - 1]. From the plot, it is evident that for the EDR2 films, the OD ranges from 0.204 to 3.04 for the dose range of 0 to 700 cGy with the linear response up to 500 cGy. In contrast, an EBT film shows a narrow OD range of 0.42 to 1.02.
Post-irradiation color growth of EBT films
Many authors (Chu et al,[3] Muench et al[4] and Meigooni et al[5]) have reported the color density growth of previous models of Gafchromic films. To determine the time dependence of change in color density of irradiated EBT film, five pieces of EBT film were irradiated to the doses of 19.6 cGy, 98 cGy, 147 cGy, 196 cGy and 294 cGy respectively. The optical densities of these films were measured at different times after irradiation for about 40 h. The scanner value measured at different times after irradiation for above-mentioned doses is plotted in [Figure - 2]. From the plot, it is evident that the scanner value decreases as a function of time. The percentage of increase in density between 2 and 40 h after irradiation is tabulated in [Table - 1].
Film uniformity
Previous models of radiochromic films have showed a large variation in uniformity, particularly along the coating direction.[6],[7],[8] To study the spatial uniformity of the EBT film, the film was irradiated by 3 x 3 cm field at 12 different places (a 3 x 4 matrix) for equal dose of 150 cGy. The film was scanned with the filter 24 h after irradiation. All other experimental films were scanned 2 h after irradiation, but this film was scanned 24 h after irradiation to allow for the maximum color growth. The irradiated film and relative values of OD, normalized to minimum OD measured, corresponding to irradiated positions are shown in [Figure - 3]. From the figure, it was evident that the film uniformity varied up to 8.5% between two regions of the film.
Effect of scanning direction on sensitometry of EBT film
To study the effect of scanning direction on scanner value, the irradiated EBT films were scanned along both portrait and landscape directions. The mean scanner value was plotted against dose for both the scans [Figure - 4]. It is evident from the plot that the overall scanner values for landscape direction scanning seem to be higher than that of portrait scanning for the same film and dose. The explanation given by the manufacturer for this is the anisotropic scattering of transmitted light by the needle-like active components, which are aligned in a direction of smaller dimension of the film.
IMRT dose distribution verification
The IMRT plans generated by Brainscan planning system were verified using EDR2 and EBT films. For the axial dose comparison, hybrid plans of patient plans were generated on scanned images of Med-Tec IMRT phantom. The EDR2 and EBT films were sandwiched together and irradiated simultaneously for plan verification. The intensity pattern verifications were performed by irradiating the films in a plane perpendicular to the beam axis at a depth of 1.5 cm in a solid water phantom. Scanned films were calibrated using their respective calibration data set. The fluence and dose distribution patterns corresponding to the film positions in the phantoms were derived from respective planning systems. The film and calculated dose pattern are co-registered for analysis. Using the template registration methods in RIT113 software, the plan-film fiducial points and their respective template points were registered unambiguously. The qualitative comparisons like overlaid isodoses [Figure - 5]a and b and profiles [Figure - 6]a and b and quantitative comparisons like Gamma index with a tolerance of 3% dose difference and 3 mm distance agreement were performed to check the consistency between calculated and measured dose distributions [Table - 2].
Results and Conclusions | |  |
The Gafchromic EBT film has a better sensitivity when compared to the previous versions of Gafchromic films and that falls in the clinical dose range. Though EBT films respond with the narrow OD range, due to the high resolving power of VXR16 digitizing system, over 65,000 gray shades can be identified and about 50,000 fall within the OD of 1; the small variations in OD produced by the EBT films are interpretable. There is a significant growth in density even 2 h after irradiation; generally the percentage of growth in density increases with time and dose. The dose measurements of films scanned at different times after irradiation will affect the accuracy of measurements because of this differential density growth.
One of the following three methods can eliminate the effect of density growth on the measurement:
a) By applying appropriate growth correction factors for individual doses
b) By maintaining the same time difference between the film irradiation and scanning for both calibration and experimental films
c) Creating a number of calibration data sets by scanning the calibration films at different times after irradiation so that appropriate calibration data set can be used depending on the scanning of experimental film after irradiation.
In our study, we have maintained the same time difference between the film irradiation and scanning for both calibration and experimental film to eliminate the effect of density growth.
For the same film and dose, the scanner value varies with the scanning direction of the EBT film. According to the explanation given by the manufacturer, this is due to the anisotropic transmission of light by the active components present in the coating of EBT film. The active component of the EBT film is a tiny needle-like structure. Because of their shape, they tend to align along their long axes parallel to the coating direction of the film. Normally, the coatings are performed in a direction parallel to the smaller (8") side of the film. Because of the shape and alignment of the active components, relatively more light is scattered normal to the direction of coating than parallel to the direction of coating. It is necessary to scan all the calibration and experiment film in a same orientation and it is necessary to have a track on film orientation if it has been cut into pieces. Since the portrait scanning shows higher sensitivity of the film, it is recommended to perform portrait scanning.
In 15 dose comparisons done with both EDR2 and EBT films, EDR2 films show good and consistent agreement between measured and calculated dose distributions. The quantitative analysis between the calculated and measured dose distribution was evaluated using Gamma function. The mean Gamma values and percentage of pixels exceeding the tolerance are shown in [Table - 2]. From the tabulated data, it was evident that EDR2 films showed consistent and good agreement between the calculated and measured dose distribution, whereas with the EBT films, the results obtained were very inconsistent. For two specific studies (Study 7 and 13), the results obtained with EBT films were comparable with the results obtained with EDR2 films. But for other studies, the mean gamma and the percentage of pixels exceeding the tolerance values varied considerably when compared to the results obtained using EDR2 films. Better results were obtained for those two studies because of relatively smaller area of dose distribution (4.5 x 4.5 cm) and smaller number of pixels analyzed when compared to other studies. This strongly suggests that the variation in the spatial uniformity of the EBT film led to the large variation in the dose verification. For dose distributions of smaller area, better results can be achieved because over a small area, the film uniformity does not vary heavily. The measurement accuracy for the conventional large field dose distributions can be improved using double-irradiation method by delivering unknown dose followed by the delivery of known uniform dose to get the uniformity correction matrix of the experimental film. But it is a complex task; moreover, at present no software tool is commercially available to apply the uniformity correction matrix on the experimental film and perform the dose comparisons. So the inference from our study is that EBT film can be used for small field IMRT verification rather than the large field IMRT under present conditions.
References | |  |
1. | Burch SE, Keafortt KJ, Trueblood JH, Sheils WC, Yeo JI, Yang CK. A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering. Med Phys 1997;24:775-83. |
2. | Niroomand A, Blackwell CR, Coursey BM, Gall KP, Galvin JM, McLaughlin WL, et al . Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55. Med Phys 1998;25:2093-115. |
3. | Chu RD, Van Dyke G, Lewis DF, O'Hara KP, Buckland BR, Dinelle F. Gaf-chromic dosimetry media: A new high dose rate thin film routine dosimeter and dose mapping tool. Radiat Phys Chem 1990;35:767-73. |
4. | Meigooni AS, Sanders MF, Ibbott GS, Szeglin SR. Dosimetry Characteristics of an improved radiochromic film. Med Phys 1996;23:1883-8. [PUBMED] |
5. | Muench PJ, Meigooni AS, Nath R, McLaughlin WL. Photo energy dependence of the sensitivity of radiochromic film and comparison with silver halide film and LiF TLD's used for brachytherapy dosimetry. Med Phys 1991;18:769-75. [PUBMED] |
6. | Mishra V, Williamson JF, Li Z, Meigooni AS. Radiochromic film densitometry for Ir-192: Accuracy and precision. Med Phys 1993;20:865. |
7. | Meigooni AS, Klevin EE, Mishra V, Low DA, Purdy JA. Gafchromic film dosimetry for megavoltage X-ray beams. Med Phys 1993;20:888. |
8. | Zhu Y, Kirov AS, Mishra V, Meigooni AS, Williamson JF. Quantitative evaluation of radiochromic film response for two-dimensional dosimetry. Med Phys 1995;22:995. |
Figures
[Figure - 1], [Figure - 2], [Figure - 3], [Figure - 4], [Figure - 5], [Figure - 6] Tables
[Table - 1], [Table - 2]
This article has been cited by | 1 |
Evaluation of patient-specific IMRT quality assurance and point dose measurement for complex head and neck and brain cancer using Gafchromic EBT3 film |
|
| Varsha R Gedam, Anirudh Pradhan | | Radiation Protection Dosimetry. 2022; | | [Pubmed] | [DOI] | | 2 |
Report of AAPM Task Group 235 Radiochromic Film Dosimetry: An Update to TG-55 |
|
| Azam Niroomand-Rad, Sou-Tung Chiu-Tsao, Michael P. Grams, David F. Lewis, Christopher G. Soares, Leo J. Van Battum, Indra J. Das, Samuel Trichter, Michael W. Kissick, Guerda Massillon-JL, Paola E. Alvarez, Maria F. Chan | | Medical Physics. 2020; 47(12): 5986 | | [Pubmed] | [DOI] | | 3 |
Characterization of dose delivery in a hard X-ray irradiation facility |
|
| Ki-Man Lee,So-Ra Kim,Eun-Hee Kim | | Journal of Nuclear Science and Technology. 2012; 49(6): 655 | | [Pubmed] | [DOI] | | 4 |
Dosimetry of gamma chamber blood irradiator using Gafchromic EBT film |
|
| S. Vandana, V.S. Shaiju, S.D. Sharma, S. Mhatre, S. Shinde, G. Chourasiya, Y.S. Mayya | | Applied Radiation and Isotopes. 2011; 69(1): 130 | | [VIEW] | [DOI] | | 5 |
Measurement of relative output factors for the 8 and 4 mm collimators of Leksell Gamma Knife Perfexion by film dosimetry |
|
| Novotny, J., Bhatnagar, J.P., Quader, M.A., Bednarz, G., Lunsford, L.D., Huq, M.S. | | Medical Physics. 2009; 36(5): 1768-1774 | | [Pubmed] | | 6 |
Measurement of relative output factors for the 8 and 4 mm collimators of Leksell Gamma Knife Perfexion by film dosimetry |
|
| Josef Novotny,Jagdish P. Bhatnagar,Mubina A. Quader,Greg Bednarz,L. Dade Lunsford,M. Saiful Huq | | Medical Physics. 2009; 36(5): 1768 | | [Pubmed] | [DOI] | | 7 |
The use of new GAFCHROMIC[sup ®] EBT film for [sup 125]I seed dosimetry in Solid Water[sup ®] phantom |
|
| Sou-Tung Chiu-Tsao, David Medich, John Munro | | Medical Physics. 2008; 35(8): 3787 | | [VIEW] | [DOI] | |
|
 |
|
|
|
|